<> "The repository administrator has not yet configured an RDF license."^^ . <> . . . "Myocardial T1-mapping at 3T using saturation-recovery: reference values, precision and comparison with MOLLI"^^ . "Background: Myocardial T1-mapping recently emerged as a promising quantitative method for non-invasive tissue characterization in numerous cardiomyopathies. Commonly performed with an inversion-recovery (IR) magnetization preparation at 1.5T, the application at 3T has gained due to increased quantification precision. Alternatively, saturation-recovery (SR) T1-mapping has recently been introduced at 1.5T for improved accuracy. Thus, the purpose of this study is to investigate the robustness and precision of SR T1-mapping at 3T and to establish accurate reference values for native T1-times and extracellular volume fraction (ECV) of healthy myocardium. Methods: Balanced Steady-State Free-Precession (bSSFP) Saturation-Pulse Prepared Heart-rate independent Inversion-REcovery (SAPPHIRE) and Saturation-recovery Single-SHot Acquisition (SASHA) T1-mapping were compared with the Modified Look-Locker inversion recovery (MOLLI) sequence at 3T. Accuracy and precision were studied in phantom. Native and post-contrast T1-times and regional ECV were determined in 20 healthy subjects (10 men, 27 ± 5 years). Subjective image quality, susceptibility artifact rating, in-vivo precision and reproducibility were analyzed. Results: SR T1-mapping showed <4 % deviation from the spin-echo reference in phantom in the range of T1 = 100–2300 ms. The average quality and artifact scores of the T1-mapping methods were: MOLLI:3.4/3.6, SAPPHIRE:3.1/3.4, SASHA:2.9/3.2; (1: poor - 4: excellent/1: strong - 4: none). SAPPHIRE and SASHA yielded significantly higher T1-times (SAPPHIRE: 1578 ± 42 ms, SASHA: 1523 ± 46 ms), in-vivo T1-time variation (SAPPHIRE: 60.1 ± 8.7 ms, SASHA: 70.0 ± 9.3 ms) and lower ECV-values (SAPPHIRE: 0.20 ± 0.02, SASHA: 0.21 ± 0.03) compared with MOLLI (T1: 1181 ± 47 ms, ECV: 0.26 ± 0.03, Precision: 53.7 ± 8.1 ms). No significant difference was found in the inter-subject variability of T1-times or ECV-values (T1: p = 0.90, ECV: p = 0.78), the observer agreement (inter: p > 0.19; intra: p > 0.09) or consistency (inter: p > 0.07; intra: p > 0.17) between the three methods. Conclusions: Saturation-recovery T1-mapping at 3T yields higher accuracy, comparable inter-subject, inter- and intra-observer variability and less than 30 % precision-loss compared to MOLLI."^^ . "2016" . . . "18" . "84" . . "BioMed Central"^^ . . . "Journal of Cardiovascular Magnetic Resonance"^^ . . . "1532429X" . . . . . . . . . . . . . . . . . . . . . . . . . . "Lothar R."^^ . "Schad"^^ . "Lothar R. Schad"^^ . . "Johannes"^^ . "Budjan"^^ . "Johannes Budjan"^^ . . "Uwe"^^ . "Mattler"^^ . "Uwe Mattler"^^ . . "Dirk"^^ . "Loßnitzer"^^ . "Dirk Loßnitzer"^^ . . "Sebastian"^^ . "Weingärtner"^^ . "Sebastian Weingärtner"^^ . . "Nadja M."^^ . "Meßner"^^ . "Nadja M. Meßner"^^ . . "Frank G."^^ . "Zöllner"^^ . "Frank G. Zöllner"^^ . . "Theano"^^ . "Papavassiliu"^^ . "Theano Papavassiliu"^^ . . . . . . "Myocardial T1-mapping at 3T using saturation-recovery: reference values, precision and comparison with MOLLI (PDF)"^^ . . . "12968_2016_Article_302.pdf"^^ . . . "Myocardial T1-mapping at 3T using saturation-recovery: reference values, precision and comparison with MOLLI (Other)"^^ . . . . . . "lightbox.jpg"^^ . . . "Myocardial T1-mapping at 3T using saturation-recovery: reference values, precision and comparison with MOLLI (Other)"^^ . . . . . . "preview.jpg"^^ . . . "Myocardial T1-mapping at 3T using saturation-recovery: reference values, precision and comparison with MOLLI (Other)"^^ . . . . . . "medium.jpg"^^ . . . "Myocardial T1-mapping at 3T using saturation-recovery: reference values, precision and comparison with MOLLI (Other)"^^ . . . . . . "small.jpg"^^ . . . "Myocardial T1-mapping at 3T using saturation-recovery: reference values, precision and comparison with MOLLI (Other)"^^ . . . . . . "indexcodes.txt"^^ . . "HTML Summary of #22187 \n\nMyocardial T1-mapping at 3T using saturation-recovery: reference values, precision and comparison with MOLLI\n\n" . "text/html" . . . "610 Medizin"@de . "610 Medical sciences Medicine"@en . .