eprintid: 23763 rev_number: 13 eprint_status: archive userid: 3447 dir: disk0/00/02/37/63 datestamp: 2017-12-05 13:21:03 lastmod: 2017-12-07 08:15:05 status_changed: 2017-12-05 13:21:03 type: doctoralThesis metadata_visibility: show creators_name: Tric, Mircea title: Optical in-line biosensor for long-term continuous glucose monitoring and control in cell culture subjects: 570 divisions: 160100 adv_faculty: af-14 abstract: Monitoring important process variables such as glucose in real-time is a major goal of bioprocess engineering, because it allows process control, which is not only essential for product quality and yield, but also important for the documentation and understanding of the production process and therefore relates to risk management [1]. This thesis deals with the development, thorough characterization and application of a disposable, optical in-line biosensor for monitoring and control of glucose in suspension cell culture. The in-line sensor, developed in this thesis, utilizes a commercially available oxygen sensor, that is coated with a crosslinked glucose oxidase (GOD) enzyme layer. The sensitivity of the sensor was tuned by the addition of a hydrophilic perforated diffusion membrane, to customize the dynamic range in order to meet the desired specifications. The biosensor was modelled in order to gain crucial insights into the internal concentration profile of the enzyme deactivating by-product hydrogen peroxide. The one-dimensional biosensor model revealed that the turnover rate of the enzyme GOD plays a crucial role for the functional stability of the biosensor in combination with the internal hydrogen peroxide accumulation. This insight was utilized to optimize the glucose biosensor for long-term continuous glucose monitoring over typical cell culture durations. A comprehensive biosensor characterization was performed to study the applicability and limitations of the developed biosensor for cell culture. Hereby, it was demonstrated that the sensor is sterilisable with beta, gamma and UV irradiation and is only subject to minor cross sensitivity to oxygen in combination with a reference oxygen sensor. The presented optical biosensor provides information in real-time and was therefore used in combination with a reference oxygen sensor to control the glucose level continuously in CHO cell culture with an automated feeding systems. It was found that the sialylation of the hyperglycosylated erythropoietin analog Darbepoetin alfa, could be significantly increased through continuous glucose feeding by retaining a high glucose level during the production phase of the cell culture. Therefore, the developed biosensor provides a valuable tool for optimizing culture conditions in biotechnological applications. date: 2017 id_scheme: DOI id_number: 10.11588/heidok.00023763 ppn_swb: 1657360822 own_urn: urn:nbn:de:bsz:16-heidok-237637 date_accepted: 2017-10-24 advisor: HASH(0x564e1c3c8420) language: eng bibsort: TRICMIRCEAOPTICALINL2017 full_text_status: public place_of_pub: Heidelberg citation: Tric, Mircea (2017) Optical in-line biosensor for long-term continuous glucose monitoring and control in cell culture. [Dissertation] document_url: https://archiv.ub.uni-heidelberg.de/volltextserver/23763/1/Dissertation%20Mircea%20Tric.pdf