title: Measurement of isocenter accuracy and image distortion in magnetic resonance-guided radiotherapy creator: Dorsch, Stefan subject: ddc-500 subject: 500 Natural sciences and mathematics subject: ddc-530 subject: 530 Physics subject: ddc-600 subject: 600 Technology (Applied sciences) description: Magnetic resonance (MR)-guided radiotherapy involves complex irradiation- and imaging devices (MR-Linacs), as well as complex treatment procedures. To assure accurate patient treatments, both have to be tested for proper functioning. Here, new methods to simultane-ously measure the isocenter alignment accuracy and the geometric image distortions of clin-ical MR-Linacs were developed. A new phantom was designed, which includes a polymer dosimetry gel-container to visualize the beams from a star shot and to identify the radiation isocenter position in MR as well as a regular grid used to visualize the MR image distortions in 3D. It was found that the gel can be evaluated immediately after irradiation with a geo-metric accuracy comparable to that of radiochromic films. The method was applied on a clinical 0.35 T MR-Linac and the isocenter alignment accuracy in 3D was found to be (0.8 ± 0.9) mm. The spatial image distortions after machine-specific correction were (0.60 ± 0.28) mm and 99.82% of the 1330 evaluated control points within a 140 mm sphere had devi-ations below 1.5 mm. Finally, a 3D printing materials and printing technique, compatible with the polymer dosimetry gel, was identified for future designs of anthropomorphic phantoms to be used in end-to-end tests in MR-guided radiotherapy. date: 2020 type: Dissertation type: info:eu-repo/semantics/doctoralThesis type: NonPeerReviewed format: application/pdf identifier: https://archiv.ub.uni-heidelberg.de/volltextserverhttps://archiv.ub.uni-heidelberg.de/volltextserver/27896/1/Stefan%20Dorsch_PhD-Thesis.pdf identifier: DOI:10.11588/heidok.00027896 identifier: urn:nbn:de:bsz:16-heidok-278967 identifier: Dorsch, Stefan (2020) Measurement of isocenter accuracy and image distortion in magnetic resonance-guided radiotherapy. [Dissertation] relation: https://archiv.ub.uni-heidelberg.de/volltextserver/27896/ rights: info:eu-repo/semantics/openAccess rights: http://archiv.ub.uni-heidelberg.de/volltextserver/help/license_urhg.html language: eng