eprintid: 36763 rev_number: 13 eprint_status: archive userid: 9100 dir: disk0/00/03/67/63 datestamp: 2025-06-26 06:24:40 lastmod: 2025-06-26 06:25:13 status_changed: 2025-06-26 06:24:40 type: doctoralThesis metadata_visibility: show creators_name: Faller, Debora title: Development of the Diamond windows - Paris-Edingburgh cell (D-PEC) for ambient and high temperature experiments subjects: ddc-550 divisions: i-121000 adv_faculty: af-12 keywords: Paris-Edinburgh press, diamond window cterms_swd: Hoher Druck abstract: In this thesis, a gasket with two diamond windows was constructed for a Paris- Edinburgh cell (D-PEC) to enable in-situ measurements. The sample pressure was determined by means of the fluorescence peak shift of ruby and the Raman peak shift of quartz. The calibration of the sample pressure against the oil pres- sure of the D-PEC at ambient temperature was carried out up to a pressure of 5 GPa. In addition, a gasket equipped with an oven and two diamond windows was constructed, allowing to conduct experiments at temperatures up to 1000 K. This HT gasket was calibrated up to pressures of 6 GPa using fluorescence measurements of ruby and samarium-doped strontium borate and Raman mea- surements of quartz. In addition, the D-PEC was irradiated with swift heavy ions for the first time at 272 K and 0.15 GPa. Amorphization of irradiated quartz and apatite was observed. date: 2025 id_scheme: DOI id_number: 10.11588/heidok.00036763 own_urn: urn:nbn:de:bsz:16-heidok-367636 date_accepted: 2025-05-20 advisor: HASH(0x55caa0dd0378) language: eng bibsort: FALLERDEBODEVELOPMEN full_text_status: public place_of_pub: Heidelberg citation: Faller, Debora (2025) Development of the Diamond windows - Paris-Edingburgh cell (D-PEC) for ambient and high temperature experiments. [Dissertation] document_url: https://archiv.ub.uni-heidelberg.de/volltextserver/36763/1/thesis_D.Faller.pdf