eprintid: 4893 rev_number: 8 eprint_status: archive userid: 1 dir: disk0/00/00/48/93 datestamp: 2004-09-02 10:57:30 lastmod: 2014-04-03 14:00:37 status_changed: 2012-08-14 15:12:54 type: doctoralThesis metadata_visibility: show creators_name: Razavi Khosravani Nejad, Karaneh title: Entwicklung und Optimierung von Messmethoden zur in-vivo-Bestimmung der MR-Relaxationsparameter T1 und T2 am menschlichen Herzen title_en: Development and Optimization of Methods for in-vivo Determination of the MR parameters T1 and T2 of the Human Heart ispublished: pub subjects: 530 divisions: 850300 adv_faculty: af-13 keywords: Myokardiale Perfusion , Myokardiale OxygenierungMyocardial Perfusion , Myocardial Oxygenation cterms_swd: NMR-Tomographie abstract: Die MR-Relaxationszeiten T1 und T2 können zur Charakterisierung der myokardialen Mikrozirkulation benutzt werden. Die T1-Zeit erlaubt die Berechnung der myokardialen Perfusion, während aus der Kenntnis der T2-Zeit über den BOLD-Effekt der Status der myokardialen Oxygenierung erfasst werden kann. Im Rahmen dieser Arbeit wurden Messmethoden zur kontrastmittelfreien in-vivo-Messung der myokardialen Relaxationszeiten auf einem klinischen Tomographen implementiert und weiterentwickelt. Für die T1-Messung wurden zwei Gradientenechosequenzen basierend auf der FLASH- und der TrueFISP-Technik entwickelt. Durch die gemessenen T1-Zeiten konnte bei ersten Probandenstudien die myokardiale Perfusion mit FLASH zu (4,2 ± 1,1) ml×g-1×min-1 und TrueFISP zu (5,5 ± 1,1) ml×g-1×min-1 bestimmt werden. Das im Myokard gemessene Signal-zu-Rausch-Verhältnis war bei TrueFISP mit 96,9 ± 6,7 fast dreifach höher als bei der FLASH-Technik mit 36,1 ± 3,7. Der mit der TrueFISP-Sequenz gemessene myokardiale Perfusionsanstieg zwischen Ruhe- und Stresszustand nach Gabe von Adenosin betrug 62,5%. Zur Bestimmung der T2-Zeit des Myokards wurde eine Navigator-TSE und eine HASTE-Sequenz optimiert. TSE- und HASTE-Sequenz lieferten mit (64 ± 11) ms bzw. (61 ± 2) ms fast die gleiche T2-Zeit, jedoch nur die HASTE-Sequenz erlaubt mit einer Messzeit von 20 s (zum Vergleich TSE: 20 min) eine T2-Bestimmung im Stresszustand. abstract_translated_text: MR relaxation times T1 and T2 can be used for the calculation of myocardial microcirculation. T1 allows the determination of myocardial perfusion, while BOLD-effect connects T2 with myocardial oxygenation. In the present work, techniques for in-vivo determination of myocardial relaxation time without contrast agent were developed and optimized. Fast gradient-echo-sequences based on FLASH and TrueFISP techniques were developed to determine myocardial T1. Myocardial perfusion was determined in volunteers with FLASH to (4.2 ± 1.1) ml×g-1×min-1 and with TrueFISP to (5.5 ± 1.1) ml×g-1×min-1. The signal-to-noise-ratio in myocard of the TrueFISP sequence was with 96.9 ± 6.7 almost three times higher than with FLASH 36.1 ± 3.7. The rise of myocardial perfusion between rest and adenosine-induced cardiac stress measured by TrueFISP was 62.5%. A navigator TSE and a HASTE-sequence were optimized to determine myocardial T2. The measured T2 by TSE and HASTE is (64 ± 11) ms and (61 ± 2) ms, respectively. Due to the short acquisition time of the HASTE sequence (20 s) (for comparison TSE: 20 min), T2 measurement under cardiac stress is possible. abstract_translated_lang: eng date: 2004 date_type: published id_scheme: DOI id_number: 10.11588/heidok.00004893 ppn_swb: 1643839632 own_urn: urn:nbn:de:bsz:16-opus-48934 date_accepted: 2004-07-21 advisor: HASH(0x564e1c57f190) language: ger bibsort: RAZAVIKHOSENTWICKLUN2004 full_text_status: public citation: Razavi Khosravani Nejad, Karaneh (2004) Entwicklung und Optimierung von Messmethoden zur in-vivo-Bestimmung der MR-Relaxationsparameter T1 und T2 am menschlichen Herzen. [Dissertation] document_url: https://archiv.ub.uni-heidelberg.de/volltextserver/4893/1/diss_krazavi_04.pdf