Direkt zum Inhalt
  1. Publizieren |
  2. Suche |
  3. Browsen |
  4. Neuzugänge rss |
  5. Open Access |
  6. Rechtsfragen |
  7. EnglishCookie löschen - von nun an wird die Spracheinstellung Ihres Browsers verwendet.

Repeatability and reproducibility of cerebral 23Na imaging in healthy subjects

Meyer, Melissa M. ; Haneder, Stefan ; Konstandin, Simon ; Budjan, Johannes ; Morelli, John N. ; Schad, Lothar R. ; Kerl, Hans U. ; Schoenberg, Stefan O. ; Kabbasch, Christoph

In: BMC Medical Imaging, 19 (2019), Nr. 26. S. 1-7. ISSN 1471-2342

[thumbnail of 12880_2019_Article_324.pdf]
Vorschau
PDF, Englisch - Hauptdokument
Download (1MB) | Lizenz: Creative Commons LizenzvertragRepeatability and reproducibility of cerebral 23Na imaging in healthy subjects von Meyer, Melissa M. ; Haneder, Stefan ; Konstandin, Simon ; Budjan, Johannes ; Morelli, John N. ; Schad, Lothar R. ; Kerl, Hans U. ; Schoenberg, Stefan O. ; Kabbasch, Christoph steht unter einer Creative Commons Namensnennung 4.0

Zitieren von Dokumenten: Bitte verwenden Sie für Zitate nicht die URL in der Adresszeile Ihres Webbrowsers, sondern entweder die angegebene DOI, URN oder die persistente URL, deren langfristige Verfügbarkeit wir garantieren. [mehr ...]

Abstract

Abstract Background Initial reports of 23Na magnetic resonance imaging (MRI) date back to the 1970s. However, methodological challenges of the technique hampered its widespread adoption for many years. Recent technical developments have overcome some of these limitations and have led to more optimal conditions for 23Na-MR imaging. In order to serve as a reliable tool for the assessment of clinical stroke or brain tumor patients, we investigated the repeatability and reproducibility of cerebral sodium (23Na) imaging in healthy subjects. Methods In this prospective, IRB approved study 12 consecutive healthy volunteers (8 female, age 31 ± 8.3) underwent three cerebral 23Na-MRI examinations at 3.0 T (TimTrio, Siemens Healthineers) distributed between two separate visits with an 8 day interval. For each scan a T1w MP-RAGE sequence for anatomical referencing and a 3D-density-adapted, radial GRE-sequence for 23Na-imaging were acquired using a dual-tuned (23Na/1H) head-coil. On 1 day, these scans were repeated consecutively; on the other day, the scans were performed once. 23Na-sequences were reconstructed according to the MP-RAGE sequence, allowing direct cross-referencing of ROIs. Circular ROIs were placed in predetermined anatomic regions: gray and white matter (GM, WM), head of the caudate nucleus (HCN), pons, and cerebellum. External 23Na-reference phantoms were used to calculate the tissue sodium content. Results Excellent correlation was found between repeated measurements on the same day (r2 = 0.94), as well as on a different day (r2 = 0.86). No significant differences were found based on laterality other than in the HCN (63.1 vs. 58.7 mmol/kg WW on the right (p = 0.01)). Pronounced inter-individual differences were identified in all anatomic regions. Moderate to good correlation (0.310 to 0.701) was found between the readers. Conclusion Our study has shown that intra-individual 23Na-concentrations in healthy subjects do not significantly differ after repeated scans on the same day and a pre-set time interval. This confirms the repeatability and reproducibility of cerebral 23Na-imaging. However, with manual ROI placement in predetermined anatomic landmarks, fluctuations in 23Na-concentrations can be observed.

Dokumententyp: Artikel
Titel der Zeitschrift: BMC Medical Imaging
Band: 19
Nummer: 26
Verlag: BioMed Central
Ort der Veröffentlichung: London
Erstellungsdatum: 05 Jun. 2019 11:31
Erscheinungsjahr: 2019
ISSN: 1471-2342
Seitenbereich: S. 1-7
Institute/Einrichtungen: ?? i-69200 ??
?? old-i-62700 ??
DDC-Sachgruppe: 610 Medizin
Freie Schlagwörter: 23Na (sodium) imaging, Oncologic imaging, Cerebral magnetic resonance imaging
Leitlinien | Häufige Fragen | Kontakt | Impressum |
OA-LogoDINI-Zertifikat 2013Logo der Open-Archives-Initiative